HIGH-CURRENT AND MICROSECOND DURATION PULSED HOLLOW-CATHODE LAMP EXCITED INDUCTIVELY-COUPLED PLASMA ATOMIZER ATOMIC FLUORESCENCE SPECTROMETRY
Full Text
Date
1995Author
GONG, ZB
弓振斌
YANG, PY
杨芃原
WANG, XR
王小如
HUANG, BL
黄本立
Collections
- 化学化工-已发表论文 [14469]
Abstract
ICP-AFS excited by hollow cathode lamp(HCL) with high current (15 A) and microsecond pulse (1.5 mu s) power supply has been studied. The detection limit (2 sigma) for Ca obtained with an 850 W ICP was 0.07 ng/mL, which was 40 times better than that obtained with conventionally pulsed (cp.) HCL excited at the same ICP power, and 30 times better than the best detection limit for cp. HCL, which was obtained with an 800 W ICP.