光化学荧光分析法测定头孢氨苄的研究
Determination of Cephalexin by Photochemical Fluorimetry
Abstract
在PH4.0的缓冲溶液中,头孢氨苄在紫外光照射下能发生光化学反应,形成具有荧光的产物,λ_(EX)为345nM,λ_(EM)乙为432nM。在头孢氨苄浓度为0.1~4.0μg/Ml范围内,荧光强度与浓度呈良好的线性关系,由此建立了头孢氨苄的光化学荧光测定法。方法的检出限为0.01μg/Ml,相对标准偏差为0.93%。本法简便、快速且灵敏度高,可用于尿液中头孢氨苄含量的测定。 Cephalexin, a nonFluorescent substance, is transFormed into a highly Fluorescent compound (λ_(ex)= 345 nm,λ_(em) = 432 nm) under the irradiation of ultraviolet light in pH 4.0 buFFer medium.Thus, a photochemical-Fluorimetric method For the determination of cephalexin was developed with a detection limit of 0.01 μg/ml and a relative standard deviation of 0.93%.The linear relationship between the Fluorescence intensity and the concentration of cephalexin is over the range of 0.1-0.4 μg/ml.This method is simple, rapid and sensitive, and has been sucessFully applied to the determination of cephalexin in urine.